SCC-AI Series Isolated Analog Input Modules User Guide 10 ni.com
Power Requirement
Analog power .........................................410 mW max
+15 V ...............................................13.67 mA max
–15 V ...............................................13.67 mA max
Digital power (+5 V) ..............................650 mW max
130 mA max
Physical
Figure 4. SCC-AI Dimensions
Weight ....................................................37 g (1.0 oz)
I/O connectors.........................................One 20-pin right-angle
male connector,
one 4-pin screw terminal,
removable
Screw terminal wire gauge .....................24 to 12 AWG
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth (inputs)........................42.4 V
peak
or 60 VDC,
Measurement Category I
Caution This device is rated for Measurement Category I and is intended to carry signal
voltages no greater than 42.4 V
peak
or 60 VDC. Do not use this device for connection to
signals or for measurements within Categories II, III, or IV.