Appendix D Common Questions
© National Instruments Corporation D-3 SCXI-1125 User Manual
In LabVIEW, can I use different input limits for the same SCXI-1125
channel if I repeat the channel in the SCXI channel string array?
No, the SCXI-1125 cannot dynamically change the gain settings during
scanning. Therefore, channels with similar input ranges should be grouped
together in the channel string array. Make sure that repeated channels in
different indices of the channel string array have the same input limits in
the corresponding input limits array.
In LabVIEW, can I use virtual channels with parallel mode channels
on the SCXI-1125?
Yes, virtual channels work with parallel mode operation on the SCXI-1125.
The E/M Series DAQ device must be directly connected to the module in
parallel mode operation. Also, virtual channels that use built-in CJC are
disabled and cannot be used in parallel mode.
In LabVIEW, can I use the calgnd channel string when the SCXI-1125
is in parallel mode?
Yes, you can autozero the SCXI-1125 in LabVIEW when using the module
in parallel mode.
Table D-2. Digital Signals on the SCXI-1125
Traditional
DAQ Signal
Name
DAQmx
Signal Name
SCXI Signal
Name
50-Pin
Connector
68-Pin
Connector
Direction
DIO0 P0.0 SER DAT IN 25 52 Output
DIO4 P0.4 SER DAT OUT 26 19 Input
DIO1 P0.1 DAQ D*/A 27 17 Output
DIO2 P0.2 SLOT 0 SEL* 29 49 Output
SCANCLK AI HOLD
COMP,
AI HOLD
SCAN CLK 36 46 Output
EXTSROBE* EXTSROBE SER CLK 37 45 Output
STARTSCAN AI SAMP
CLK,
AI SAMP
SYNC* 46 38 Output