OPERATING INSTRUCTIONS AND SPECIFICATIONS
NI 9217
4-Channel, 24-Bit, 100Ω RTD Analog Input Module
ni.com/manuals
DeutschFrançais
Zobrazit stránku 0
1 2 3 4 5 6 ... 27 28

Shrnutí obsahu

Strany 1 - DeutschFrançais

OPERATING INSTRUCTIONS AND SPECIFICATIONSNI 92174-Channel, 24-Bit, 100Ω RTD Analog Input Moduleni.com/manualsDeutschFrançais

Strany 2 - Safety Guidelines

NI 9217 Operating Instructions and Specifications 10 ni.comIf you are using a 4-wire RTD, connect EX to the positive lead of the RTD. If you are using

Strany 3

© National Instruments Corp. 11 NI 9217 Operating Instructions and SpecificationsWiring for High-Vibration ApplicationsIf an application is subject to

Strany 4

NI 9217 Operating Instructions and Specifications 12 ni.comSleep ModeThis module supports a low-power sleep mode. Support for sleep mode at the system

Strany 5

© National Instruments Corp. 13 NI 9217 Operating Instructions and SpecificationsSpecificationsThe following specifications are typical for the range

Strany 6

NI 9217 Operating Instructions and Specifications 14 ni.comConversion timeHigh-resolution mode... 200 ms per channel,800 ms total for al

Strany 7 - Connecting the NI 9217

© National Instruments Corp. 15 NI 9217 Operating Instructions and SpecificationsNoiseHigh-resolution mode... 0.003 °CHigh-speed mode...

Strany 8

NI 9217 Operating Instructions and Specifications 16 ni.comHigh-resolution filter response1,21 This image is provided courtesy of Linear Technology C

Strany 9

© National Instruments Corp. 17 NI 9217 Operating Instructions and SpecificationsOvervoltage protection ... ±30 V between inputsMTBF

Strany 10

NI 9217 Operating Instructions and Specifications 18 ni.comPhysical CharacteristicsIf you need to clean the module, wipe it with a dry towel.Screw-ter

Strany 11

© National Instruments Corp. 19 NI 9217 Operating Instructions and SpecificationsSafetyMaximum Voltage1 Connect only voltages that are within the foll

Strany 12 - Sleep Mode

NI 9217 Operating Instructions and Specifications 2 ni.comThis document describes how to use the National Instruments 9217 and includes specifications

Strany 13 - Specifications

NI 9217 Operating Instructions and Specifications 20 ni.comCaution Do not connect the NI 9217 to signals or use for measurements within Measurement Ca

Strany 14

© National Instruments Corp. 21 NI 9217 Operating Instructions and SpecificationsHazardous LocationsU.S. (UL) ...

Strany 15

NI 9217 Operating Instructions and Specifications 22 ni.comOperating humidity(IEC 60068-2-56)... 10 to 90% RH,noncondensing

Strany 16

© National Instruments Corp. 23 NI 9217 Operating Instructions and SpecificationsElectromagnetic CompatibilityThis product is designed to meet the req

Strany 17 - Power Requirements

NI 9217 Operating Instructions and Specifications 24 ni.comni.com/certification, search by module number or product line, and click the appropriate li

Strany 18 - Physical Characteristics

© National Instruments Corp. 25 NI 9217 Operating Instructions and SpecificationsCalibrationYo u can obtain the calibration certificate and informatio

Strany 19 - Isolation Voltages

NI 9217 Operating Instructions and Specifications 26 ni.comWhere to Go for SupportThe National Instruments Web site is your complete resource for tech

Strany 20 - Safety Standards

© National Instruments Corp. 27 NI 9217 Operating Instructions and SpecificationsKorea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 8877

Strany 21 - Environmental

© 2006–2008 National Instruments Corp. All rights reserved.374187C-01 May08National Instruments, NI, ni.com, and LabVIEW are trademarks of National In

Strany 22 - Shock and Vibration

© National Instruments Corp. 3 NI 9217 Operating Instructions and SpecificationsHot Surface This icon denotes that the component may be hot. Touching

Strany 23 - CE Compliance

NI 9217 Operating Instructions and Specifications 4 ni.comFigure 1 shows the NI 9939 connector backshell.Figure 1. NI 9939 Connector BackshellSafety

Strany 24 - Environmental Management

© National Instruments Corp. 5 NI 9217 Operating Instructions and Specificationsonly. Follow these guidelines if you are installing the NI 9217 in a p

Strany 25 - Calibration

NI 9217 Operating Instructions and Specifications 6 ni.comSpecial Conditions for Hazardous Locations Use in EuropeThis equipment has been evaluated as

Strany 26 - Where to Go for Support

© National Instruments Corp. 7 NI 9217 Operating Instructions and SpecificationsConnecting the NI 9217The NI 9217 has a 16-terminal, detachable screw-

Strany 27

NI 9217 Operating Instructions and Specifications 8 ni.comYo u can connect any combination of 3-wire or 4-wire RTDs to the NI 9217. Each NI 9217 chann

Strany 28 - 374187C-01 May08

© National Instruments Corp. 9 NI 9217 Operating Instructions and SpecificationsFigure 3. Input Circuitry for One Channel of the NI 9217 in 4-Wire Mo