© Copyright 1993, 1994 National Instruments Corporation.All Rights Reserved.SCXI-1163User Manual32-Channel Isolated Digital Output ModuleSeptember 199
About This Manual© National Instruments Corporation xi SCXI-1163 User ManualAbbreviations, acronyms, metric prefixes, mnemonics, symbols, and terms ar
IndexSCXI-1163 User Manual Index-4 © National Instruments Corporationparts locator diagramjumper configuration, 2-2SCXI-1326 terminal block, 2-13PC-LP
Index© National Instruments Corporation Index-5 SCXI-1163 User Manualregister selection and writeprocedure, 5-2 to 5-3register writes, 5-1 to 5-3SCXIb
IndexSCXI-1163 User Manual Index-6 © National Instruments Corporationselecting serial mode with jumperW6, 2-6serial data timing diagram, 2-20serial I/
Index© National Instruments Corporation Index-7 SCXI-1163 User Manualequivalents for rear signalconnector, 3-4pin assignments, 3-2signal descriptions,
About This ManualSCXI-1163 User Manual xii © National Instruments CorporationCustomer CommunicationNational Instruments wants to receive your comments
© National Instruments Corporation 1-1 SCXI-1163 User ManualChapter 1Introduction
Introduction Chapter 1SCXI-1163 User Manual 1-2 © National Instruments CorporationWith the SCXI-1163, you can use the SCXI chassis as a controller in
Chapter 1 Introduction© National Instruments Corporation 1-3 SCXI-1163 User ManualYou can also use the SCXI-1163, combined with the NB Series data acq
Introduction Chapter 1SCXI-1163 User Manual 1-4 © National Instruments Corporation• ERNI Components, Inc. (part number 033-273)The mating connector f
© National Instruments Corporation 2-1 SCXI-1163 User ManualChapter 2Configuration and Installation
Chapter 2 Configuration and Installation© National Instruments Corporation 2-3 SCXI-1163 User ManualThe SCXI-1163 can accept data in two forms–serial
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-4 © National Instruments CorporationDo not operate damaged equipment. The safety prot
National Instruments Corporate Headquarters6504 Bridge Point ParkwayAustin, TX 78730-5039(512) 794-0100Technical support fax: (800) 328-2203(512) 794-
Chapter 2 Configuration and Installation© National Instruments Corporation 2-5 SCXI-1163 User Manual- Jumper W5 carries the SCXIbus MISO line, after b
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-6 © National Instruments CorporationPosition PAR disconnects SERDATOUT from the SCXIbu
Chapter 2 Configuration and Installation© National Instruments Corporation 2-7 SCXI-1163 User ManualTable 2-1. Configuration of Jumpers W2, W3, and W
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-8 © National Instruments CorporationHardware InstallationYou can install the SCXI-1163
Chapter 2 Configuration and Installation© National Instruments Corporation 2-9 SCXI-1163 User ManualFront ConnectorFigure 2-2 shows the pin assignment
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-10 © National Instruments CorporationFront Connector Signal DescriptionsPin Signal Nam
Chapter 2 Configuration and Installation© National Instruments Corporation 2-11 SCXI-1163 User ManualPin Signal Name Description (continued)A1, B1, A2
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-12 © National Instruments CorporationWhen using the terminal block with high common-mo
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-14 © National Instruments CorporationSCXI-1326 InstallationTo connect the terminal blo
Limited WarrantyThe SCXI-1163 is warranted against defects in materials and workmanship for a period of one year from the date ofshipment, as evidence
Chapter 2 Configuration and Installation© National Instruments Corporation 2-15 SCXI-1163 User Manual1 23 45 67 89 1011 1213 14
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-16 © National Instruments CorporationRear Signal Connector Signal Descriptions, Serial
Chapter 2 Configuration and Installation© National Instruments Corporation 2-17 SCXI-1163 User ManualAll other pins are not connected.The signals on t
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-18 © National Instruments CorporationWhen you configure the SCXI-1163 for an MIO-type
Chapter 2 Configuration and Installation© National Instruments Corporation 2-19 SCXI-1163 User ManualTiming Requirements and Communication ProtocolCom
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-20 © National Instruments CorporationTo write the 16-bit slot-select number to Slot 0,
Chapter 2 Configuration and Installation© National Instruments Corporation 2-21 SCXI-1163 User ManualTo write to the Address Handler, follow these ste
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-22 © National Instruments CorporationTo write to the Data Register, follow these steps
Chapter 2 Configuration and Installation© National Instruments Corporation 2-23 SCXI-1163 User ManualTo read from the Module ID Register, follow these
Configuration and Installation Chapter 2SCXI-1163 User Manual 2-24 © National Instruments CorporationTo read from a Status Register, follow these step
© National Instruments Corporation v SCXI-1163 User ManualContents
© National Instruments Corporation 3-1 SCXI-1163 User ManualChapter 3Theory of Operation
Theory of Operation Chapter 3SCXI-1163 User Manual 3-2 © National Instruments CorporationINTR*D*/ACHSGNDCHSGNDV+V++5 VMOSI A1B1A2B2A3B3A4B4A5B5A6B6A7
Chapter 3 Theory of Operation© National Instruments Corporation 3-3 SCXI-1163 User ManualSCXIbus Connector Signal DescriptionsPin Signal Name Descript
Theory of Operation Chapter 3SCXI-1163 User Manual 3-4 © National Instruments CorporationTable 3-1. SCXIbus Equivalents for Rear Signal ConnectorRear
Chapter 3 Theory of Operation© National Instruments Corporation 3-5 SCXI-1163 User ManualThis circuitry is divided into an SCXI digital interface sect
Theory of Operation Chapter 3SCXI-1163 User Manual 3-6 © National Instruments CorporationThe register circuitry section consists of the Data Register,
Chapter 3 Theory of Operation© National Instruments Corporation 3-7 SCXI-1163 User Manualoutput channels and the Vcc and ground connections. Table 3-
© National Instruments Corporation 4-1 SCXI-1163 User ManualChapter 4Register Descriptions
Register Descriptions Chapter 4SCXI-1163 User Manual 4-2 © National Instruments CorporationAddress HandlerYou must write to the Address Handler at the
Chapter 4 Register Descriptions© National Instruments Corporation 4-3 SCXI-1163 User ManualModule ID RegisterThe Module ID register contains the 4-byt
ContentsSCXI-1163 User Manual vi © National Instruments CorporationChapter 4Register Descriptions...
Register Descriptions Chapter 4SCXI-1163 User Manual 4-4 © National Instruments CorporationData RegisterThe Data Register contains 32 bits that contro
Chapter 4 Register Descriptions© National Instruments Corporation 4-5 SCXI-1163 User ManualStatus RegistersThe Status Registers are 1-bit read-only re
Register Descriptions Chapter 4SCXI-1163 User Manual 4-6 © National Instruments CorporationParking RegisterThis register is a dummy register that you
Chapter 4 Register Descriptions© National Instruments Corporation 4-7 SCXI-1163 User ManualSlot 0 Register DescriptionsSlot 0 has three registers–the
Register Descriptions Chapter 4SCXI-1163 User Manual 4-8 © National Instruments CorporationSlot-Select RegisterThe Slot-Select Register contains 16 bi
© National Instruments Corporation 5-1 SCXI-1163 User ManualChapter 5Programming
Programming Chapter 5SCXI-1163 User Manual 5-2 © National Instruments CorporationTable 5-1. SCXIbus to SCXI-1163 Rear Signal Connector to Data Acquis
Chapter 5 Programming© National Instruments Corporation 5-3 SCXI-1163 User Manual5. If you are writing to a Data Register, you must first write addres
Programming Chapter 5SCXI-1163 User Manual 5-4 © National Instruments CorporationExamplesThis section describes how to program the SCXI-1163, either a
Chapter 5 Programming© National Instruments Corporation 5-5 SCXI-1163 User ManualExample 3You have a multiple-chassis system with an SCXI-1163 in Slot
Contents© National Instruments Corporation vii SCXI-1163 User ManualAppendix ESCXI-1163 Cabling...
© National Instruments Corporation A-1 SCXI-1163 User ManualAppendix ASpecifications
Specifications Appendix ASCXI-1163 User Manual A-2 © National Instruments CorporationOperating EnvironmentTemperature 0° to 50° CRelative humidity 5%
© National Instruments Corporation B-1 SCXI-1163 User ManualAppendix BRear Signal Connector
Rear Signal Connector Appendix BSCXI-1163 User Manual B-2 © National Instruments Corporation1 23 45 67 89 1011 1213 1415 1617
Appendix B Rear Signal Connector© National Instruments Corporation B-3 SCXI-1163 User ManualRear Signal Connector Signal Descriptions, Serial Configur
Rear Signal Connector Appendix BSCXI-1163 User Manual B-4 © National Instruments CorporationAll other pins are not connected.See the Timing Requiremen
© National Instruments Corporation C-1 SCXI-1163 User ManualAppendix CSCXIbus Connector
SCXIbus Connector Appendix CSCXI-1163 User Manual C-2 © National Instruments CorporationINTR*D*/ACHSGNDCHSGNDV+V++5 VMOSI A1B1A2B2A3B3A4B4A5B5A6B6A7B
Appendix C SCXIbus Connector© National Instruments Corporation C-3 SCXI-1163 User ManualSCXIbus Connector Signal DescriptionsPin Signal Name Descripti
© National Instruments Corporation D-1 SCXI-1163 User ManualAppendix DSCXI-1163 Front Connector
ContentsSCXI-1163 User Manual viii © National Instruments CorporationFiguresFigure 1-1. SCXI-1163 Digital Output Module...
SCXI-1163 Front Connector Appendix DSCXI-1163 User Manual D-2 © National Instruments CorporationPin NumberSignal NameColumn A B CSignal Name
Appendix D SCXI-1163 Front Connector© National Instruments Corporation D-3 SCXI-1163 User ManualFront Connector Signal DescriptionsPin Signal Name Des
SCXI-1163 Front Connector Appendix DSCXI-1163 User Manual D-4 © National Instruments CorporationPin Signal Name Description (continued)A1, B1, A2, Out
© National Instruments Corporation E-1 SCXI-1163 User ManualAppendix ESCXI-1163 Cabling
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-2 © National Instruments Corporation• The SCXI-1340 has an extra male breakout connector for use w
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-3 SCXI-1163 User ManualSCXI-1340 InstallationFollow these steps to install the SCXI-1
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-4 © National Instruments CorporationSCXI-1341 Lab-NB, Lab-PC, or Lab-PC+ and SCXI-1344Lab-LC Cable
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-5 SCXI-1163 User ManualSCXI-1341 and SCXI-1344 InstallationFollow these steps to inst
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-6 © National Instruments CorporationTable E-3. SCXI-1342 Pin TranslationsPC-LPM-16 Pin PC-LPM-16
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-7 SCXI-1163 User ManualDIO-96, AT-MIO-16D, and AT-MIO-64F-5 BoardConnectionTo use you
© National Instruments Corporation ix SCXI-1163 User ManualAbout This Manual
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-8 © National Instruments CorporationTable E-4. SCXI-1163 and DIO-96, AT-MIO-16D, and AT-MIO-64F-5
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-9 SCXI-1163 User ManualTable E-4. SCXI-1163 and DIO-96, AT-MIO-16D, and AT-MIO-64F-5
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-10 © National Instruments CorporationAfter step 1, the order of these steps is not critical; howev
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-11 SCXI-1163 User ManualTable E-5. SCXI-1348 Pin TranslationsDIO-32F Pin DIO-32F Sig
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-12 © National Instruments CorporationSCXI-1348 InstallationFollow these steps to install the SCXI-
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-13 SCXI-1163 User ManualStep 3SCXI-1180 Breakout ConnectorRibbon Cable to Front Panel
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-14 © National Instruments CorporationFront Panel ConnectorFront PanelFront Threaded StripRibbon Ca
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-15 SCXI-1163 User Manual50-Pin ConnectorGrounding ScrewThumbscrew CutoutStep 1Step 2I
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-16 © National Instruments CorporationSCXI-1351 InstallationFollow these steps to install the SCXI-
Appendix E SCXI-1163 Cabling© National Instruments Corporation E-17 SCXI-1163 User Manual4. Connect another ribbon cable to the chassis extender conne
About This ManualSCXI-1163 User Manual x © National Instruments Corporation• The Glossary contains an alphabetical list and description of terms used
SCXI-1163 Cabling Appendix ESCXI-1163 User Manual E-18 © National Instruments CorporationTable E-6. SCXI-1343 Pin Connections (Continued)Rear SignalC
© National Instruments Corporation F-1 SCXI-1163 User ManualAppendix FCustomer Communication
Technical Support Form___________________________________________________ Photocopy this form and update it each time you make changes to your softwar
SCXI-1163 Hardware Configuration Form
Documentation Comment Form
© National Instruments Corporation Glossary-1 SCXI-1163 User ManualGlossary___________________________________________________ Prefix Meaning Valuep-
GlossarySCXI-1163 User Manual Glossary-2 © National Instruments CorporationMSB most significant bitRAM random-access memoryRESET* Resetrms root mean s
© National Instruments Corporation Index-1 SCXI-1163 User ManualIndex
IndexSCXI-1163 User Manual Index-2 © National Instruments CorporationDD*/A signalSCXIbus connector, 3-3, C-3writing to Data Register, 3-6DAQD*/A signa
Index© National Instruments Corporation Index-3 SCXI-1163 User ManualSCXI-1351 one-slot cableextender, E-16unpacking the SCXI-1163 module, 1-4interfac
Komentáře k této Příručce